Products Personal dosimeters Thermoluminescent Dosimetry TLD chips

TLD Dosimeters

Thermoluminescent Dosimetry

A choice of thermoluminescent dosimeters (TLD-dosimeters), models: TLD‑3, DTL‑02, DVNG‑M, DTU‑1, DTU‑2, Finger Ring Type G, Finger Ring Type BG, EYE‑D.

To be used together with DOZA‑TLD reader.

TLD dosimeters with DOZA‑TLD reader and DVG software form a thermoluminescent dosimetry monitoring system (TLD‑system).

О Комплексе

Operating principle / measuring cycle: 1) TLD detectors (TLD chips) inside TLD dosimeter collect energy of ionizing radiation proportional to dose of exposure. 2) Been heated the TLD chips release collected energy in the form of light emission. 3) Photomultiplier tube (PMT) located in TLD reader detects the emitted light and converts it into electrical signal.

Personal TLD-dosimeter models:


TLD-3


Measuring of: personal dose equivalent Hр(10) of photon radiation
Detector: LiF:Mg,Ti; LiF:Mg,Cu,P
Number of detectors (TLD-chips) in dosimeter assembly: 3
Measuring range: 20 μSv ÷ 10 Sv
Energy range: 0,015 ÷ 18,0 MeV
Size: 62×21×12,4 mm
Weight: 16 g

DTL-02


Measuring of: personal dose equivalent Hр(10) of photon radiation
Detector: LiF:Мg,Ti 
Number of detectors (TLD-chips) in dosimeter assembly: 3
Measuring range: 20 μSv ÷ 10 Sv
Energy range: 0,015 ÷ 18,0 MeV
Size: 62×25×14 mm
Weight: 25 g


DTU-1


Measuring of:
 - personal dose equivalent Hр(10) of photon radiation
 - ambient dose equivalent H*(10) of photon radiation
Detector: LiF:Мg,Ti 
Number of detectors (TLD-chips) in dosimeter assembly: 2
Measuring range: 20 μSv ÷ 10 Sv 
Energy range:
 - 0,015 ÷ 18,0 MeV for personal dose equivalent Hр(10)
 - 0,03 ÷ 3,0 MeV for ambient dose equivalent H*(10)
Size: 42×28×18 mm
Weight: 8 g


DTU-2


Measuring of:
 - personal dose equivalent Hр(10) of photon radiation
 - ambient dose equivalent H*(10) of photon radiation
Detector: Al2O3
Number of detectors (TLD-chips) in dosimeter assembly: 2
Measuring range: 20 μSv ÷ 0,5 Sv
Energy range:
 - 0,08 ÷ 6,0 MeV for personal dose equivalent Hр(10)
 - 0,08 ÷ 3,0 MeV for ambient dose equivalent H*(10)
Size: 42×28×18 mm
Weight: 8 g


DVNG-M



Measuring of: personal dose equivalent Hр(10) of photon and neutron radiation
Detector: 6LiF:Mg,Ti; 7LiF:Mg,Ti
Number of detectors (TLD-chips) in dosimeter assembly: 4 (2 - 6LiF:Mg,Ti; 2 - 7LiF:Mg,Ti)
Measuring range:
- 100 μSv ÷ 2 Sv for neutron radiation
 - 20 μSv ÷ 10 Sv for photon radiation
Energy range:
 - 0,025 эВ ÷ 20,0 MeV for neutron radiation
 - 0,015 ÷ 18,0 MeV for photon radiation 
Size: 48×40×20 mm
Weight: 30 g


FINGER RING TYPE G



Measuring of: personal dose equivalent Hp(0.07) of photon radiation for skin
Detector: LiF:Мg,Ti or LiF:Mg,Cu,P
Number of detectors (TLD-chips) in dosimeter assembly: 1
Measuring range: 20 μSv ÷ 10 Sv
Energy range: 0,01 ÷ 6,0 MeV
Size: 91×12×3 mm
Weight: 1 g

FINGER RING TYPE BG


Measuring of: personal dose equivalent Hp(0.07) of photon and beta radiation for skin
Detector: LiF:Mg,Cu,P
Number of detectors (TLD-chips) in dosimeter assembly: 1
Measuring range: 20 μSv ÷ 10 Sv
Energy range: 
 - 0,0076 ÷ 6,0 MeV for photon radiation 
 - 0,05 ÷ 3,0 MeV for beta radiation 
Size: 91×12×3 mm
Weight: 1 g


EYE-D



Measuring of: personal dose equivalent Hp(3) of photon and beta radiation for eye lens
Detector: LiF:Мg,Ti or LiF:Mg,Cu,P
Number of detectors (TLD-chips) in dosimeter assembly: 1
Measuring range: 20 μSv ÷ 10 Sv
Energy range: 
 - 0,015 ÷ 18,0 MeV for photon radiation
 - 0,7 ÷ 3,0 MeV for beta radiation 
Size: 71×16×8 mm
Weight: 3 g