Monitoring of radioactive aerosols in the work area
Purpose
Continuous measurement of volumetric activity of radioactive aerosols due to artificial alpha-emitting nuclides
Sound/light alarm in the event that the measured values of volumetric activity exceed the preset threshold levels
Features
Stationary installation at the working area air monitoring points
Connection to the pump unit BN-01
Or another external sampling device
Service interface USB for diagnostics, adjustment and calibration
Physical characteristics
Measurement range of volumetric activity: 1,0 ÷ 2,0·10
5
Bq/m
3
Energy range of alpha particles detection: 3,0 ÷ 8,0 MeV
Sensitivity coefficient for particles within energy range based on
239
Pu: Not less than 0.02
Intrinsic background of alarm unit: not more than 1,0·10
-1
Bq/m
3
Air flow rate through the filter: 20 ÷ 40 l/min
Measurement range of air volume pumped throughthe filter: 10 ÷ 100 000 l/min
Power supply: 220 V, 50 Hz
Electrical characteristics
Power consumption: 50 V・A
Mechanical characteristics
Overall dimensions: 350×156×482 mm
Weight: not more than 6 kg
Back to the Section