Purpose:
- monitoring of alpha and beta contamination of hands, feet (footwear) and personal working/protection clothes; alarm signals for preset thresholds;
Features:
- warning light “READY” (ready for measurement);
- visible and audible alarms “CONTAMINATION” (presence of contamination in excess of the threshold) and “CLEAN” (contamination is absent);
- individual threshold setting for each measurement channel;
- automatic compensation for background gamma radiation and contamination of detector units;
- indication of the contamination level for each detector unit from a system keyboard;
- input of threshold settings from a system keyboard;
- automatic self-test capability for detectors and measurement channels;
- indication of contaminated part(s) of the body on the panel with mnemonic image of the body;
- network capability through communication interfaces RS-485 and RS-232;
- availability of different models allows the user to purchase a system equipped with necessary number of measurement channels (from 15 to 20), optional remote detector units, detectors for monitoring of dorsal sides of hands, sensors èëè detectors for monitoring of head and the built-in computer.
Controlled areas of the body:
Channel
|
Detector unit
|
Monitored part of the body
|
Comment
|
1
|
BDZB
|
Face
|
Stationary (built-in) measurement channels
|
2
|
BDZB
|
Chest
|
3
|
BDZB
|
Palm of left hand
|
4
|
BDZB
|
Palm of right hand
|
5
|
BDZB
|
Abdomen
|
6
|
BDZB
|
Left knee
|
7
|
BDZB
|
Right knee
|
8
|
BDZB
|
Back of the head
|
9
|
BDZB
|
Shoulder-blade, neck
|
10
|
BDZB
|
Back
|
11
|
BDZB
|
Buttocks
|
12
|
BDZB
|
Sural area of the right leg
|
13
|
BDZB
|
Sural area of the left leg
|
14
|
BDZB
|
Foot of the left leg
|
15
|
BDZB
|
Foot of the right leg
|
16
|
BDZB
|
Dorsal side of the left hand
|
17
|
BDZB
|
Dorsal side of the right hand
|
18
|
BDZB
|
Upper area of the head
|
19
|
BDZB
|
Measurement of beta contamination
|
Remote measurement channels
|
20
|
BDZA
|
Measurement of alpha contamination
|
21
|
BDZB
|
Measurement of background gamma radiation
|
Built-in measurement channel
|
Delivery options:
Racks with two exit doors: 20, 21, 22, 23, 24, 25, 28, 29, 32, 36, 68.
Racks with one exit door: 40, 41, 42, 44, 48, 49, 52, 54, 56.
Model
|
Configuration
|
20, 40
|
Maximum configuration:
- stand with 18 stationary BDZB detector units;
- one remote detector unit BDZB-05 for beta radiation;
- one remote detector unit BDZA-96 for alpha radiation;
- BOI-04.
|
21, 41
|
- stand with 15 stationary BDZB detector units for beta contamination measurements;
- one remote channel for beta radiation;
- one remote channel for alpha radiation.
|
22, 42
|
- stand with 15 stationary BDZB detector units for beta contamination measurements (analog of RZB-98).
|
24, 44
|
- stand with 15 stationary BDZB detector units for beta contamination measurements;
- one remote channel for beta radiation;
- BOI-04.
|
28, 48
|
- stand with 18 stationary BDZB detector units;
- one remote channel for beta radiation (BDZB-05).
|
29, 49
|
- stand with 18 stationary BDZB detector units;
- one remote channel for alpha radiation (BDZA-96).
|
32, 52
|
- stand with 18 stationary BDZB detector units for beta contamination measurements
|
34, 54
|
- stand with 15 stationary BDZB detector units (equipped with magnets of traveling column for pitching conditions);
- marine model.
|
36, 56
|
- stand with 17 stationary BDZB detector units (15 + 2 units for dorsal sides of hands).
|
68
|
- stand with 36 stationary BDZB detector units;
- built-in computer;
- one remote channel for beta radiation;
- one remote channel for alpha radiation;
- BOI-04.
|