Energy range of gamma and X-ray radiations
|
0.015 ÷ 2.5 MeV
|
| Measurement range of dose Hp(10) |
0.01 mSv ÷ 25 mSv
|
|
Limits for basic relative measurement error of personal dose equivalent
|
less than ±25 %
|
|
Self-discharge of the dosimeter during 1 hour does not exceed:
|
|
|
0.5 μSv
|
- under high temperature conditions (up to +50 °Ñ)
|
1 μSv
|
- under low temperature conditions (down to minus 50°Ñ)
|
1 μSv
|
- under high humidity (up to 98 % at 35 °C)
|
1 μSv
|
|
Power supply:
|
|
|
4 elements, type CR 1025
|
- reading and control device KSU-01
|
mains 220V, 50Hz; built-in batteries
|
|
Autonomous continuous operation time without charging of built-in batteries (reading of 40 dosimeters twice a day)
|
not less than 30 days
|
|
Computer interface
|
RS-232/USB 2.0
|
|
Operation temperature range:
|
|
|
minus 40 °C ÷ +50 °C
|
- reading and control device
|
+10 °C ÷ +45 °C
|
|
Overall dimensions, weight:
|
|
|
diam. 18×126 mm, 0.05 kg
|
- reading and control device
|
235×232×125 mm, 2.8 kg
|